Model 428 User’s Manual REMOTE COMMUNICATIONS
Bristol Instruments, Inc. 48
Measurement Instruction Functions
:POWer?
Returns a power reading in either Watts or dBm as specified by the
:UNIT:POWer function.
For a SCALar query, one numeric value is returned. For an ARRay
query, an integer value indicating the number of entries, followed by
comma separated numerical values in fixed point notation.
(e.g., 2, -12.67, -15.49)
:READ:ARR:POW?
:FETC:SCAL:POW?
Returns the instrument's internal temperature in degrees Celsius (ºC)
and pressure in millimeters of mercury (mm Hg). The response is
independent of the choice of SCALar or ARRay.
Two numerical values with units separated by a comma (e.g., 28.5 C,
740 MMHG).
:MEAS:ARR:ENV?
:FETC:ARR:ENV?
Returns an input laser reading in units of THz.
A single numerical value (e.g., 193.67357) for SCALar.
An integer value indicating the number of entries, followed by an array of
comma separated numerical values (e.g. 2, 193.67357, 193.77482) for
ARRay.
:FETC:ARR:FREQ?
:MEAS:ARR:FREQ?
Returns an input laser wavelength value in units of nm.
A single numerical value (e.g., 1531.5877) for SCALar.
An integer value indicating the number of entries, followed by an array of
comma separated numerical values (e.g. 2, 1531.5877, 1532.7832) for
ARRay.
:FETC:SCAL:WAV?
:READ:ARR:WAV?