Model 428 User’s Manual REMOTE COMMUNICATIONS
Bristol Instruments, Inc. 53
:THREShold {REL | ABS}
Sets the threshold type to either relative (REL) or absolute (ABS).
Queries the factory set wavelength calibration coefficient. Used to
properly scale the x axis of the spectrum. See Appendix F.
Queries the wavelength limit state.
:WLIMit [:STATe] {ON | OFF}
Turns the wavelength limiting to ON or OFF. If ON, the channel
identification and wavelength calculations are limited to
Queries the wavelength start or stop wavelength.
A single numerical value in nanometer units.
:CALC2:WLIM:STAR?
:CALC2:WLIM:STOP?
Sets the threshold type to either relative (REL) or absolute (ABS).
:CALC2:WLIM:STAR 1490
:CALC2:WLIM:STOP 1580