EasyManua.ls Logo

Brooks 5850 S - Section 3 Operation

Default Icon
52 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Installation and Operation Manual
X-TMF-5800S-MFC-eng
PN 541-C-051-AAG
November, 2008
Models 5800-S
18
Table 2-3: Electrical interfacing. Smart TMF cable
sub D (25p)
female
pin no.
sub D (9p)
female
pin no.
1. 6 brown Setpoint return (-)
2. 10 white 0 (1)-5 Vdc Flow signal output
3. 9 grey/pink (TTL) Open collector alarm output
4. 2 red/blue 0 (4)-20 mA Flow signal output
5. 13 red +15 Vcc to +28 Vdc Power supply
6. 14 blue -15 Vdc Power supply (if required)
7. 3 violet 0 (4)-20 mA Setpiont input (+)
8. 5 grey 0 (1)-5 Vdc Setpiont input (+)
9. 12 7 5 black Power supply common
10. 8 pink Flow signal output common
11. 4 yellow +5 Vdc ref. output
12. 7 green valve override input
11 shielding Chassis
14. 2 3 green/white Optional RS232C-RxD or RS485-A-
15. 3 2 blue/white Optional RS232C-TxD or RS485-A+
Read-out side
sub D (15p)
male
pin no.
Sub-D (15 c.)
sub D (15p)
female
pin no.
Colour Function Computer side
Section 2: Installation
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com