5.3 Parameter: Group 1
Mode 1 Code number
Value: 11
Mode 2 Range allocation for analog output
Default value: 20 mm/s
Value Equals
10 0 ... 10 mm/s
20 0 ... 20 mm/s
50 0 ... 50 mm/s
100 0 ... 100 mm/s
Switching of measuring ranges
Switching from a smaller range to a larger range:
The defined limits values are retained.
Switching from a larger range to a smaller range:
As long as the defined limit values are larger than the new full scale they will
be converted to the new full scale. If the defined limit values are lower than
the new full scale they will be retained.
The automatic change in limit values will be signalled by flashing of the
respective alarm limit LEDs (LIM 1 and LIM 2). To reset this flashing the limit
value settings must first be changed, i.e. limit values lower than the full scale
must be set and the instrument switched off and then on again for the
change to take effect.
Mode 3 Limit value LIM 1
Default value: 4.5 mm/s
The adjustable range of the limit value is from 0 to the full scale value of the
selected range. The resolution of the steps in the adjustment is in every case
dependent on the selected range.
Range Resolution
0 ... 10 mm/s 0,1 mm/s
0 ... 20 mm/s 0,1 mm/s
0 ... 50 mm/s 0,2 mm/s (x,0 - x,2 - x,5 - x,7)
0 ... 100 mm/s 1 mm/s