VIBROCONTROL 920 Instrument versions
Page 30 / 33
VC920E / C102833.002 / V 13
9 Instrument versions
Version < 2.8
Filter settings
The value for the lower filter frequency amounts to 3 Hz resp. 10 Hz
Conduct at switch-on
During the self-test after switch-on or return of power, the OK relay and limit
relays will be in the de-energized condition.
Version < 3.0
Full scale changes
If the full scale value
- Parameter group 1 Mode 2-
changes, the defined limit values
- LIM 1: Group 1 Mode 3 and
- LIM 2: Group 1 Mode 4
are retained.
Analog output test
Selecting the analog output test function
- Parameter group 2 Mode 7 / 4 -
switched an AC voltage of 5 V and 195 Hz to the diagnostic output.