Bruker Daltonik GmbH Operation
6 READY:
a) Permanently lit: Ready for acquiring spectra, i.e. any interlock is closed; N
2
pressure is correct, sufficient vacuum, HV present, etc.
b) OFF: During the docking procedure.
7 WARM UP:
a) Permanent lit: Warm-up phase of the laser device (ClassIIIb/ ClassIV Laser
Product)
in progress or the system is not ready for measurement, e.g., the
docking procedure in still in progress, HV is off, etc. It takes some time, even
when all parameters comply with the requirements. This state is reflected
simultaneously on the PC screen providing details via flexControl.
b) Flash: A subordinated malfunction has occurred, which must be verified by the
user (less N2 pressure, interlock open).
8 ERROR:
c) Permanent lit: malfunction, requires intervention of the user or a service
technician; the tool tips feature on the PC shows details (flexControl).
d) Flashing: (combined with Target Error): Service mode or firmware update.
9 MAINS:
Permanent lit, when the main circuit breaker is on.
ultraflex III User Manual, Version 1.0 46