R The solvent you wish to use is part of the solvent library. See Chapter6.3 "Editing
a solvent", page53
u Navigate to the
Solvent Loading
dialog according to the navigation path.
ð The display shows the dialog box
Solvent Loading
.
u Tap the drop-down list next to Line 1.
ð The display shows a drop-down list with selectable solvents.
u Select the solvent which is connected to solvent line 1.
ð The solvent for Line 1 is assigned.
ð The drop-down list closes.
u Repeat the solvent selection for each line.
u Activate the checkbox next to the line you wish to prime.
u Tap the button [Auto Prime].
u Wait till priming finished.
u Tap the button [Close].
ð All lines are assigned with solvents.
ð The dialog box closes.
5.8 Calibrating the solvent level sensor
Navigation path
Tools Calibration and Defaults
Precondition:
R The solvent level sensor is not covered by solvent.
u Navigate to the
Calibration
dialog according to the navigation path.
ð The display shows the dialog box
Calibration
.
u Tap [Zero] for the related solvent line.
u Tap the button [Close].
ð The dialog box closes.