5 | Installation
Büchi Labortechnik AG
28/102
Operation Manual Pure Chromatography Instruments
5.9 Installing the exhaust (ELSD only)
u Press the exhaust gas hose (2) on the
exhaust (1).
5.10 Assembling the waste bottle
NOTICE
Waste bottle on top of the instrument
Waste bottle on top of the instrument can cause property damages.
u Make sure that there is a decline between the connection on the instrument and
the waste bottle.
u Put the waste line into the waste bottle.
5.11 Installing the waste level sensor
Precondition:
R Make sure that the instrument is not
connected to the power supply.
u Attach the waste level sensor to the
instrument. Connections see
Chapter3.2.4 "Connections on gradient
valve", page17
u Calibrate the waste level sensor. See
Chapter5.12 "Calibrating the waste
level sensor", page29