3 | Product description
Büchi Labortechnik AG
20/102
Operation Manual Pure Chromatography Instruments
Pure
C-810
Pure
C-815
Pure
C-830
Pure
C-835
Pure
C-850
Pressure range - - 0-300 bar 0-300 bar 0-300 bar
Flow rate - - 0 - 100 ml/
min
0 - 100 ml/
min
0 - 100 ml/
min
Flow rate accuracy - - < 2 % < 2 % < 2 %
Functional principle - - self-priming self-priming self-priming
Specification - - 2 pistons
parallel ar-
ranged
2 pistons
parallel ar-
ranged
2 pistons
parallel ar-
ranged
UV Detector
Pure
C-810
Pure
C-815
Pure
C-830
Pure
C-835
Pure
C-850
UV - Vis wavelength
range
200 - 800
nm
200 - 800
nm
200 - 800
nm
200 - 800
nm
200 - 800
nm
Light sources Halogen /
Deuterium
Halogen /
Deuterium
Halogen /
Deuterium
Halogen /
Deuterium
Halogen /
Deuterium
Detector DAD DAD DAD DAD DAD
DAD scan entire
range
entire
range
entire
range
entire
range
entire
range
ELSD Detector
Pure
C-810
Pure
C-815
Pure
C-830
Pure
C-835
Pure
C-850
Light output laser - 0.3 mW - 1 mW 1 mW
Pressure carrier air - ~0.8 bar - ~0.8 bar ~0.8 bar
Pressure spray air - 3 - 3.5 bar - 3 - 3.5 bar 3 - 3.5 bar
Flow rate air - 2 - 2.5 L/
min
- 2 - 2.5 L/
min
2 - 2.5 L/
min
3.5.2 Ambient conditions
For indoor use only.
Max. altitude above sea level 2000 m
Ambient temperature 5‒40°C (25°C)
No maximum performance above 25°C
Maximum relative humidity 80% for temperatures up to 31°C
Storage temperature max. 45 °C