6 | Operation
Büchi Labortechnik AG
54/102
Operation Manual Pure Chromatography Instruments
u Select solvent group for the defined solvent.
u Tap the button [OK].
ð The dialog box closes.
ð The solvent is added.
u Tap the button [Close].
ð The dialog box
Solvent Definition
closes.
6.3.2 Deleting a solvent
Navigation path
Tools Solvent Definition
u Navigate to the
Solvent Definition
dialog according to the navigation path.
ð The display shows the dialog box
Solvent Definition
.
u Select the solvent you wish to delete.
u Tap the button [Delete Solvent].
u Answer the secure question with [Yes].
ð The solvent is deleted.
6.4 Tasks during a separation
6.4.1 Introducing a sample into the prep system
u Connect the syringe at the sample loop
port.