EasyManua.ls Logo

Burkert multiCELL 8619 - Page 34

Burkert multiCELL 8619
40 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
32
Adjustment and commissioning
8.9.4. Calibrating a redox sensor
Modify the default calibration limits before calibrating your sensor in the menu "Parameters" -> "Mx:pH/
ORP" -> "Calibration limits" -> "Offset ORP".
In order not to interrupt the process, activate the HOLD function in the menu "Calibration -> System ->
Hold". The Hold mode is automatically deactivated when the multiCELL restarts after a power interruption,
if the Hold mode was activated at the moment of the power cut-off.
Before each calibration, correctly clean the electrode with a suitable product.
Calibrate the redox sensor using a 1-point procedure.
The calibration may fail:
a possible "warning" message indicates either an error in the buffer solution or the ageing of the probe.
a possible "error" message indicates that the probe must be replaced.
The 1-point calibration procedure is used for a quick calibration by adjusting the zero of the measurement graph
with a buffer solution with a known oxidation reduction potential.
Immerse the clean probe in the redox solution:
the multiCELL displays the measured potential
difference of the solution
Enter the potential difference of the redox
solution (indicated on the bottle).
Calibration ORP calibrationMx:pH/ORP 1st point
OK
F4
1st point
465.0 mV
475.0 mV
ORP Calibr. result
Offset:-55.60 mV
The multiCELL displays the calibration result.
1)
Type 8619
english

Related product manuals