LySonix 3000
®
with PulseSelect™ Operation Manual
40-OM-0002 Page 7 of 43 Rev B
WARNING:
PLUGGING GENERATOR UNIT INTO SOCKET WHICH SUPPLIES
IMPROPER VOLTAGE MAY CAUSE GENERATOR TO MALFUNCTION OR
CREATE A SHOCK OR FIRE HAZARD. BE CERTAIN LINE INPUT
VOLTAGE SELECTOR SWITCH IS PROPERLY ADJUSTED.
WARNING:
PROPER SYSTEM GROUNDING CANNOT BE INSURED UNLESS UNIT IS
CONNECTED TO A PROPERLY WIRED HOSPITAL GRADE OUTLET.
CAUTION: DO NOT PLACE GENERATOR ON TOWEL, FOAM OR OTHER SOFT
SURFACE SINCE THE MATERIAL MAY BLOCK AIR VENTS LOCATED ON
THE BOTTOM OF THE GENERATOR. BLOCKING THESE VENTS MAY
CAUSE GENERATOR TO OVERHEAT AND MALFUNCTION OR CREATE A
SHOCK HAZARD.
CAUTION: SYSTEM TEST SHOULD ALWAYS BE DONE IN ADVANCE OF PREPPING
PATIENT FOR SURGERY TO MINIMIZE RISK TO PATIENT IN CASE OF
SYSTEM MALFUNCTION.
WARNING:
IF PROBE IS CRACKED, DAMAGED, OR IN NEED OF REPLACEMENT,
SURFACE TEMPERATURE CAN BECOME VERY HOT, SO THAT PROBE
MAY PRESENT A BURN HAZARD TO PATIENT OR CLINICIAN, IF
TOUCHED. EXERCISE CARE IN TESTING AND HANDLING OF PROBES
AT ALL TIMES. PROBES SHOULD BE TESTED FOR TEMPERATURE RISE
AT VARIOUS INTERVALS DURING PROCEDURE TO PREVENT BURNING
OF PATIENT DUE TO DAMAGED OR CRACKED PROBE. DO NOT TOUCH
PROBE WHILE IT IS ACTIVATED.
WARNING:
DO NOT HOLD PROBE TIP UP TO EAR. THE OPERATING FREQUENCY
OF THE SYSTEM IS ABOVE THE RANGE OF HUMAN HEARING, BUT
DOES EMIT ACOUSTIC ENERGY. DO NOT ACTIVATE THE PROBE IN
OPEN AIR WITHIN 2 FEET OF THE EARS OF OPERATING ROOM STAFF
OR PATIENT.
CAUTION: DURING CLEANING BE CERTAIN TO CLEAR DEBRIS FROM ALL
HOLLOW PROBE AND HANDPIECE INTERNAL PASSAGES BY BRUSHING.
FAILURE TO DO SO MAY HINDER STERILIZATION OF UNITS DURING
AUTOCLAVING.
CAUTION: DO NOT USE ULTRASONIC CLEANERS TO CLEAN HANDPIECE CABLES
OR PROBES. USE MANUAL CLEANING TECHNIQUES ONLY.