LySonix 3000
®
with PulseSelect™ Operation Manual
40-OM-0002 Page 8 of 43 Rev B
CAUTION: BEFORE USING LOOSE PACKING MATERIALS, SUCH AS FOAM
PELLETS, SHREDDED PAPER, OR EXCELSIOR, BE SURE TO WRAP THE
COMPONENT(S) SEPARATELY IN PLASTIC BAGS OR FILM OR OTHER
PROTECTIVE WRAPPING.
Electromagnetic Compatibility Guidance (in accordance with EN/IEC 60601-1-2:2001)
CAUTION:
THE LYSONIX 3000
®
WITH PULSESELECT
™
IS CONSIDERED MEDICAL
ELECTRICAL EQUIPMENT. MEDICAL ELECTRICAL EQUIPMENT NEEDS
SPECIAL PRECAUTIONS REGARDING ELECTROMAGNETIC
COMPATIBILITY (EMC) AND NEEDS TO BE INSTALLED AND PUT INTO
SERVICE ACCORDING TO THE EMC INFORMATION PROVIDED IN THIS
OPERATOR’S MANUAL.
CAUTION:
PORTABLE AND MOBILE RF COMMUNICATION EQUIPMENT CAN
EFFECT MEDICAL ELECTRICAL EQUIPMENT.
WARNING
THE USE OF ACCESSORIES, TRANSDUCERS AND CABLES OTHER THAN
THOSE SPECIFIED, WITH THE EXCEPTION OF TRANSDUCERS AND
CABLES SOLD BY THE MANUFACTURER OF THE LYSONIX 3000
®
WITH
PULSESELECT
™
AS REPLACEMENT PARTS FOR INTERNAL
COMPONENTS, MAY RESULT IN INCREASED EMISSIONS OR
DECREASED IMMUNITY OF THE LYSONIX 3000
®
WITH PULSESELECT
™
.
WARNING:
THE LYSONIX 3000
®
WITH PULSESELECT
™
SHOULD NOT BE USED
ADJACENT TO OR STACKED WITH OTHER EQUIPMENT AND THAT IF
ADJACENT OR STACKED USE IS NECESSARY, THE LYSONIX 3000
®
WITH PULSESELECT
™
SHOULD BE OBSERVED TO VERIFY NORMAL
OPERATION IN THE CONFIGURATION IN WHICH IT WILL BE USED.
List of cables, transducers and accessories:
Item Cable
Length
Type
Ultrasonic Transducer NA NA
Transducer cable 3.8m shielded 4-conductor
AC Main Power cord 3.6m unshielded 3-conducter
Foot Switch Cable 4.4m unshielded 2-conductor
Pump Umbilical Cable 1.5m unshielded 2-conductor