Scanner
®
2000 microEFM Section 2
Remove the plug from the conduit opening in the top of the Scanner 2000 enclosure, route the turbine
VLJQDOFDEOHWKURXJKWKHRSHQLQJDQGFRQQHFWLWWRWKHPDLQFLUFXLWERDUG$ZLULQJGLDJUDPIRUWKHWXU-
bine input is provided in )LJXUHSDJH)RUKD]DUGRXVDUHDVUHYLHZ+D]DUGRXV$UHD,QVWDOODWLRQV
page
Flow
Manifold
Static pressure input
(manifold equalizer valve
must remain open)
H
L
RTD assembly
10 pipe diameters
upstream
5 pipe diameters
downstream
Figure 2.7—Remote-mount installation in an AGA-7 turbine meter run
,QVWDOOWKH57'DVVHPEO\LQWKHWKHUPRZHOO5HPRYHWKHSOXJIURPWKHRWKHUFRQGXLWRSHQLQJLQWKHWRS
of the Scanner 2000 enclosure, route the RTD assembly cable through the conduit opening in the top of
WKH6FDQQHUDQGFRQQHFWLWWRWKHPDLQFLUFXLWERDUG$ZLULQJGLDJUDPIRUWKH57'DVVHPEO\LV
provided in )LJXUHSDJH)RUKD]DUGRXVDUHDVUHYLHZ+D]DUGRXV$UHD,QVWDOODWLRQVSDJH
=HURWKHVWDWLFSUHVVXUHDQGUHFDOLEUDWHWKHVWDWLFSUHVVXUHLIUHTXLUHG6HHWKH0RG:RU;3UR6RIWZDUH
8VHU0DQXDO3DUW1R$IRUFRPSOHWHLQVWUXFWLRQV6HHDOVR=HUR2IIVHW6WDWLF3UHVVXUHRU
'LIIHUHQWLDO3UHVVXUHSDJHDQG6WDWLF3UHVVXUH&DOLEUDWLRQDQG9HUL¿FDWLRQSDJH
CAUTION Do not put the Scanner into operation until the valves are positioned properly so that
SUHVVXUHLVVXSSOLHGWRERWKVLGHVRIWKH097)RULQVWUXFWLRQVRQSURSHUYDOYHSRVLWLRQV
see 3ODFLQJWKH6FDQQHULQWR2SHUDWLRQSDJH58.