51
Scanner
®
2000 microEFM Section 2
Measuring Compensated Liquid via a Turbine Meter
Best Practices
7KH6FDQQHUPLFUR()0FDOFXODWHVFRPSHQVDWHGOLTXLGÀRZWKURXJKDWXUELQHPHWHULQDFFRUGDQFHZLWK
$3,0306&KDQGWKHPHDVXUHPHQWSULQFLSOHVXSRQZKLFKWKH$*$VWDQGDUGLVEDVHG7KHXVHU
supplies a linear or multi-point calibration factor, and the instrument performs the required compensation
FDOFXODWLRQVEDVHGRQWKH57'LQSXW
For optimum performance, ensure that the turbine and Scanner 2000 installation complies with the industry
UHFRPPHQGDWLRQVOLVWHGEHORZ
 ,QVWDOOWKHWXUELQHÀRZPHWHULQWKHPHWHUUXQVXFKWKDWWKHUHDUHDWOHDVWQRPLQDOSLSHGLDPHWHUVXS-
VWUHDPDQG¿YHQRPLQDOSLSHGLDPHWHUVGRZQVWUHDPRIWKHPHWHU%RWKLQOHWDQGRXWOHWSLSHVKRXOGEHRI
WKHVDPHQRPLQDOVL]HDVWKHPHWHU
 6WUDLJKWHQLQJYDQHVDUHUHFRPPHQGHGIRUHOLPLQDWLQJVZLUOFRQGLWLRQV,IXVHGWKH\VKRXOGEHLQVWDOOHG
¿YHSLSHGLDPHWHUVXSVWUHDPRIWKHPHWHU
Installation Procedure—Direct Mount to a Turbine Meter (CSA Compliant)
A Scanner 2000 without the MVT bottomworks can be mounted directly to a liquid turbine meter for
PHDVXULQJOLTXLG)LJXUH$SLSHDGDSWHUDQGXQLRQDUHDWWDFKHGWRWKH6FDQQHUDOORZLQJDGLUHFW
FRQQHFWLRQWRWKHWXUELQHPHWHU
:$51,1* +$=$5'286$5($86(7KH6FDQQHULVFHUWL¿HGIRUKD]DUGRXVDUHDXVHRQO\ZKHQ
installed in accordance with applicable standards and local wiring practices. Carefully
review +D]DUGRXV$UHD,QVWDOODWLRQVSDJH27WRGHWHUPLQHVSHFL¿FLQVWDOODWLRQUHTXLUH-
PHQWVFDEOHJODQGVFRQGXLWVHDOVVLJQDOFDEOH57'HWF
Flow
Adapter/union
(CSA-approved)
RTD assembly
10 pipe diameters
upstream
5 pipe diameters
downstream
Figure 2.12—Direct-mount installation for use with a Barton 7000 Series meter