14
CAUTION: When using the dual ranging feature of the 748, do not attempt to use a
division value in the low range that is too small to provide adequate signal strength for
stable operation of the instrument. Most scale installations will allow the smaller division
to be 1/2 of the larger division value when the dual range feature is used.
The inherent sensitivity and capabilities of the 748 will allow you to select combinations that are
beyond the practical limits of today’s load cells for dependable, stable performance.
The appearance of such a situation will be instability in the lower range, while the higher range will
operate as it should.
Press to show the current setting. Press to
disable or to enable. Press to save the displayed
setting. If dual ranging is disabled (0 entered) advance to
the
INTERVAL SETTING, otherwise continue: (NOTE: If dual ranging is selected
and decimal point locations for both Intervals are not the same, the truck storage
feature will be inhibited.)
Press to show the current setting. Press
through to enter the new Interval value and press
to save it or to save the current value displayed.
Values are 1 through 9, 10, 20, 30, 40, 50, 60, 70, 80 and 90. (Interval values
other than 1, 2, 5, 10, 20 or 50 cannot be used in commercial applications.)
Press to show the current setting. Press
through to enter the desired decimal point location
and press to save it or to save the current displayed
setting. Must be the same as High Interval to be able to use Truck Storage.
0=XXXXXX 2=XXXX.XX 4=XX.XXXX
1=XXXXX.X 3=XXX.XXX
Press to show the current setting. Press
through to enter the new Low capacity value. Then
press to save it or to save the displayed value.
Press to show the current setting. Press
through to enter the new High interval value and
press to save it or to save the current displayed
setting. Available High interval values are 1 through 9, 10,
20, 30, 40, 50, 60, 70, 80 and 90. Interval values other than 1, 2, 5, 10, 20, or 50
cannot be used in commercial applications.
ENTER 0
1
dUAL r
ENTER
Lo int
Lo dP
ENTER 0
9
ENTER
0
4
ENTER
ENTER
0
ENTER
9
ENTER
Lo CAP
0
9
ENTER
ENTER
DUAL RANGING
LOW INTERVAL
LOW DECIMAL POINT
LOW CAPACITY
HIGH INTERVAL
Hi int