15
Press to show the current setting. Press
through to enter the new High capacity value. Then
press to save it or to save the displayed value.
NOTE! High Capacity must be greater than Low Capacity and High Interval
must be greater than Low Interval. ADVANCE TO
WEIGHING UNITS.
Press to show the current setting. Press
through to enter the new Interval value and press
to save it or to save the current setting. Available interval
settings are 1 through 9, 10, 20, 30, 40, 50, 60, 70, 80 and 90. Interval values
other than 1, 2, 5, 10, 20, or 50 cannot be used in commercial applications.
Press to show the current setting. Press
through to enter the new decimal point location and
press to save it oo save the current setting.
0=XXXXXX 2=XXXX.XX 4=XX.XXXX
1=XXXXX.X 3=XXX.XXX
Press to show the current setting. Press
through to enter the new Capacity value. Then press
to save it or to save the displayed setting.
NOTE! Capacity cannot exceed 999999.
Press to show the current setting. Press through
to enter the new weighing units and press
to save it or to save the current displayed setting.
Press to show the current setting. Press through
to enter the desired decimal point location and press
to save it or to save the current setting. Must be the
same as Low Decimal Point to be able to use Truck Storage.
0=XXXXXX 2=XXXX.XX 4=XX.XXXX
1=XXXXX.X 3=XXX.XXX
Hi dP
ENTER
0
HIGH DECIMAL POINT
4
ENTER
int=
Hi CAP
ENTER
0
9
ENTER
0
9
ENTER
ENTER
1=Pounds only 3=Pounds/Kilograms 4=Kilograms/Pounds 5=Tons (tn) only
(powers up and (powers up and
Calibrates Calibrates
2=Kilograms only in Pounds) in Kilograms) 6=Metric Tons (t) only
0
6
ENTER
9
ENTER
CAP=
UnitS=
1ENTER
ENTER
HIGH CAPACITY
INTERVAL SETTING
DECIMAL POINT SETTING
CAPACITY
WEIGHINGUNITS
4
0
ENTER
ENTER
dP=