your scale is used in a commercial application for buying
or selling, you may not use the “C” numbers. The “C”
numbers may be reviewed during normal operation by
pressing the key.
The indicator will return to the first calibration prompt, duAL r?. At this time, calibration can be
repeated if necessary. If calibration is complete, set the calibration switch to the OPR position. The
indicator will reset, ready for operation. NOTE: The calibration switch can be set to the OPR position
anytime during the calibration and setup procedure with all previous data retained.
FINE SPAN ADJUSTMENT
With power applied and displaying weight, set the calibration switch on the printed circuit board to
the CAL position. Then press to decrease span or to increase span. Return the switch to
the OPR position for normal operation.
CALIBRATION “C” NUMBERS OR CALDAC (cont)
CALIBRATION/SETUP COMPLETED
C4=
C3=
0
TEST
1