E. THE SPAN CALIBRATION WITH A PARTIAL LOAD
For the purpose of convenience, a partial span calibration is provided.
E.1 SET TO THE CALIBRATION MODE
The CAL switch is located underneath of a hole left of the upper case.
1.
Remove a seal plate.
2.
Slide CAL switch to the CAL position.
3. Plug into the AC outlet.
-> The display shows 'CAL" three times.
E.2 ENTRY OF A PARTIAL CALIBRATION
Enter to partial span calibration mode as below
E.3 INPUT A PARTIAL LOAD BY PERCENTAGE
In this mode, below three keys have an alternative functions respectively.
Enter a desired partial load by percentage using above three keys. Press the "ON/OFF" key to move to MODE 4.