CHAPTER-VI THE ERROR
MESSAGES
A. THE ERROR MESSAGES
These error messages will guide you to do
a
quick troubleshooting.
A.I 'Err 1*
The "Err 1" can be happen when a current zero point has shifted from the last
span calibration.
A.1.1 If a dead load of load cell has changed.
To use with a different dead load from with a last span calibration,
the span calibration must be done before use.
However, do not exceed of a dead load bigger than 20,000 counts in
calibration mode 5. A. 1.2 If the contact resistance of a load cell connector is poor.
Clean the load cell connector or replace it. A. 1.3 If the zero point of a load cell has
shifted.
Replace the load cell. A. 1.4 If the analog module has failed.
Replace the analog module.
For ordering, call for the name CAM 01.
A. 2 'Err 2"
A.2.1 The "Err 2" is not a real error, only it prompts return CAL switch to the normal position.
A. 3 'Err 10*
A.3.1 The "Err 10" means a failure of the analog module. Replace the analog module by a new
one. For ordering, call for the name CAM 01.
A. 4 "Err 11*
A.4.1 The "Err 11" means a writing error of the internal nonvolatile memory.
To recognize this error, be sure to voltages on the circuits and do a calibration
procedures.
Nevertheless, the display shows same "Err 11" replace the digital module.
For ordering, call for the name COM 01 version 2.
A.5 "Err 12*
A. 5.1 The "Err 12" warns that the scale has lost of the parameters for under
weighing regulations or has lost of the factors for a digital span calculation. To recover this, enter each condition codes again.
Refer to APPENDIX-I.
A. 6 'Err 13*
A.6.1 The "Err 13" means the soft key(s) code has(have) lost.
To recover from this(these) error(s), find the which key(s) has(have)
lost of the soft keycode(s) and then re-enter the soft key code(s) on it(them).
Refer to APPENDIX-I.