INSTALLING
Loadcell connection
Loadcell connector connect with the loadcell port of indicator back
panel.
EX+
EX-
SIG-SIG+
Load Cell
Connector
Resolution to loadcell output rate.
5V impression to loadcell
Max. loadcell output
Recommended
resolution
2 mV 1/2,000 (Max.)
5 mV 1/5,000 (Max.)
10 mV 1/10,000 (Max.)
Loadcell and connector connection
method
Reference : please refer to following loadcell color
because the wire color may be different by loadcell
production company and model.
Loadcell color table by company
Company
Connector
CAS
BONGSIN
JEANGWOO
KYOWA
INTERFACE
P.T
BLS
SHOWA
SHINKOH
TMI
TML
TFAC
HUNTLEIGH
No. 1 No. 2 No. 3 No. 4 No. 5
(EX+) (EX-) (SIG+) (SIG-) (SHIELD)
RED WHITE GREEN BLUE EXTERNAL
RED WHITE GREEN BLUE YELLOW
RED WHITE GREEN BLUE YELLOW
RED BLACK GREEN WHITE EXTERNAL
RED BLACK GREEN WHITE EXTERNAL
RED BLACK GREEN WHITE EXTERNAL
GREEN BLACK WHITE RED YELLOW
RED BLUE WHITE BLACK EXTERNAL
RED WHITE GREEN WHITE EXTERNAL
RED WHITE GREEN BLUE YELLOW
RED BLACK WHITE GREEN EXTERNAL
RED BLUE WHITE BLACK YELLOW
GREEN BLACK RED WHITE EXTERNAL