9
Service Manual
PB
Series
REV NO
:
(8) Press [TARE] key, and then “MCAPA“ message blinks.“0005” will be shown. The first
digit,”0” will blink. It means a mid-capability is “05 (calibration unit, kg or lb)
(9) Input a capability by using [ZERO] key.
(10)Press [TARE] key, and then “W-dP message blinks.“3” will be shown. The first digit,”3”
will blink. It means a weight decimal point is “3 (will display 0.000)”
(11)Input a weight decimal point by using [ZERO] key.
(12) Press [TARE] key, and then “1d “ message blinks.“0.005” will be shown. The third
digit,”0” will blink. It means a division is “0.005 (calibration unit, kg or lb)”
(13) Input a division by using [ZERO] key.
(14) Press [TARE] key, and then “dual message blinks.“1” will be shown. The third digit,”1”
will blink. It means a dual interval is disable. (0 : disable, 1 : enable)”
(15) Input a dual interval enable by using [ZERO] key.
(16) Press [ TARE] key to save the calibration factor, and C- 10 message will be shown.
- CAL METHOD (SPAN ,%)
- PROGRESSED TEST FACTS IN CAL MODE
- SETTING IS RELATED TO KEY VALUE
- SETTING FACTS IN THE OTHER CAL MODES (Gravity setting)
- SETTING FACTS IN SET MODE: INCLUDE A FUNCTION
4.1.5. Displaying Real A/D Value (C-5)
Display Raw AD
4.1.6. Percent Calibration (C-7)
(1) Pressing and holding “Calibration Switch” press [POWER] key.
After “CAL message blinks three times and shows the version of scale, it displays “CAL 1”
message.
(2) Press [ZERO] to display “CAL-7”.
(3) Press [TARE] key and then it displays “per 0 ” message. Select the percent value using the
[numeric] key. You can choose 10~90 percent. (Last digit of percent must be 0.)
(4) Press [TARE] key and then it displays “zero” message
(5) Press [TARE] key and then it displays “pspan ” message
(6) Load choice percentage weight of full weight on the platform
(7) Press [TARE] key and then it displays “CAL 7” message

Other manuals for CAS PB Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the CAS PB Series and is the answer not in the manual?

CAS PB Series Specifications

General IconGeneral
DisplayLCD
Overload ProtectionYes
Unitsg, oz
InterfaceRS-232 (optional on some models)
Auto CalibrationYes
Operating Temperature10°C to 40°C
MaterialStainless steel platform, ABS housing

Related product manuals