OPERATOR MAINTENANCE
Note: Does the car actually have gas in the tank? Have you physically verified the computer is
plugged in? Yes, we are going ask some very basic questions. Please bear with us.
Methodical verifications and the elimination of potential causes of failure are often the
quickest means of getting a unit back into operation.
Unit Performance Specifications
Verify the following items to ensure a fault in the unit is preventing it from achieving its specified
performance levels.
Standard Ventilation Spacing Requirements
Verify there is sufficient spacing around the oven for ventilation or cooling. Insufficient spacing can
adversely impact temperature performance.
• A minimum of 6 inches (152 mm) clearance is required between the sides of the oven and any
walls or partitions.
• A minimum of 12 inches (305 mm) is required between the top of the oven and overhead cover.
• A minimum of 12 inches (305 mm) is required between the back of the oven and any partitions or
objects.
Operating Temperature Range of the Unit
Verify that your constant temperature setpoint or the setpoints of your heating profile all fall within
the operating range of the oven.
• See the Temperature Range specifications for the oven on page 58.