TROUBLESHOOTING
General Information
The Intelect Advanced Therapy Systems incorporate Error Messages and Warnings to inform the user of problems or potential problems with the system,
modality or accessories. Refer to the Intelect Advanced Therapy System User Manual, part number 27429 for all Error Message Troubleshooting of the Intelect
Advanced Therapy System. The Troubleshooting information below is related to the Intelect Advanced Vacuum Electrode Module only.
PROBLEM PROBABLE CAUSE POSSIBLE REMEDY
Vacuum stops and all three vacuum mode LED’s are
flashing.
Reservoir full. Drain Reservoir. Refer to Maintenance section of this
manual.
Vacuum stops running and no LED’s are flashing. Primary Reservoir Sensor faulty. Contact dealer or DJO for service.
Vacuum not efficient. Electrodes not secure to patient. Lead Hoses and/or Electrode Cups worn or damaged. Replace Lead Hoses and/or Electrode Cups.