| Intelect
®
Mobile 2 - Service Manual 18
DEVICE DESCRIPTION
EN
WAVEFORMS (CONTINUED)
VMS™ Burst
Advice on size and type of electrodes to be used is given in device GUI "treatment guidelines" feature
VMS Burst is a symmetrical biphasic waveform delivered in a burst format.
Because the pulse is relatively short, the waveform has a low skin load,
making it suitable for applications requiring high intensities, such as muscle
strengthening protocols.
Output Mode Electrodes
Output Intensity 0-140 mA (CC) 0-140 V (CV)
Channel Mode Single, Reciprocal, Co-Contract Phase
Duration 30-400 µsec
Mode Selection CC or CV
Anti-Fatigue Off or On
Set Intensity Individual/both Channel Intensity
Setting in Reciprocal and Co-Contract modes
Cycle Time Continuous or User Defined
Burst Frequency 1-200 bps
Ramp 0-5 sec
Treatment Time 1-60 min
Available on Channels 1, 2
IRMS [mA] 0-138.56 mA
DC component No
MONOPHASIC: Monophasic Rectangular Pulsed
Advice on size and type of electrodes to be used is given in device GUI "treatment guidelines" feature
The Monophasic Rectangular Pulsed waveform is an interrupted
unidirectional current with a rectangular pulse shape.
Output Mode Electrodes
Output Intensity 0-60 mA
Phase Duration 0.1-1,000 ms
Phase Interval 5-5,000 ms
Treatment Time 1-60 min
Available on Channels 1, 2
Mode Selection CC
IRMS [mA] 0-79.60 mA
DC component Yes
MONOPHASIC: Monophasic Triangular Pulsed
Advice on size and type of electrodes to be used is given in device GUI "treatment guidelines" feature
The Monophasic Triangular Pulsed waveform is an interrupted unidirectional
current with a triangular pulse shape.
Output Mode Electrodes
Output Intensity 0-60 mA
Phase Duration 0.1-1,000 ms
Phase Interval 5-5,000 ms
Treatment Time 1-60 min
Available on Channels 1, 2
Mode Selection CC
IRMS [mA] 0-45.96 mA
DC component Yes
GALVANIC: Continuous
Advice on size and type of electrodes to be used is given in device GUI "treatment guidelines" feature
Galvanic Current is a direct current flowing in one direction only. The current
can be continuous or interrupted.
Output Mode Electrodes
Output Intensity 0-40 mA
Polarity Reversal On or Off
With Polarity Reversal On, Polarity will
change in the middle of the treatment time
Cycle Time Continuous, or User Defined
Treatment Time 1-60 min
Available on Channels 1, 2
Mode Selection CC
IRMS [mA] 0-80 mA
DC component Yes

Other manuals for Chattanooga Intelect Mobile 2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Chattanooga Intelect Mobile 2 and is the answer not in the manual?

Chattanooga Intelect Mobile 2 Specifications

General IconGeneral
BrandChattanooga
ModelIntelect Mobile 2
CategoryMedical Equipment
LanguageEnglish

Related product manuals