EasyManua.ls Logo

Chattanooga Intelect Mobile 2 - Page 63

Chattanooga Intelect Mobile 2
81 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
FINAL TEST
EN
| Intelect
®
Mobile 2 - Service Manual 63
Simulation of Electrode fault on CH2
Remove the Electrotherapy connector 2 from the
device, an advice is displayed indicating
“Electrode fault (0x71020000)”.
Check Pass/Fail the tracker sheet §8 if “Electrode
fault (0x71020000)” is displayed.
Resumption of the Electrotherapy program on
CH2
Put back the Electrotherapy connector 2 into the
device.
Press “Resume”.
You should have the CH2 status channel
“Running”.
Increase again the intensity with the knob until
the maximum, 100 mA.
Check Pass/Fail the tracker sheet §9 if the
program is resumed on CH2.
Simulation of Electrode fault on CH1
Select CH1
Remove the Electrotherapy connector 1 from the
device, an advice is displayed indicating
“Electrode fault (0x41020000)”.
Check Pass/Fail the tracker sheet §10 if
“Electrode fault (0x41020000)” is displayed.

Other manuals for Chattanooga Intelect Mobile 2

Related product manuals