FINAL TEST
EN
| Intelect
®
Mobile 2 - Service Manual 66
8.3.1. Tracker sheet for Intelect Mobile 2 COMBO Final Test Procedure
1. Serial Number (S/N) Done
2. Software Version: ⎵⎵.⎵⎵.⎵⎵ Done
Therapy
3. Both voltages values are correct after stabilization Pass Fail
4. LED on the device and G16 Test Load works and the program run without error Pass Fail
5. The display indicates “US applicator disconnected (0x00000000)” Pass Fail
6. The ultrasound program is resumed Pass Fail
7. Both LED works and channels run for 15 seconds without pause or any popped error Pass Fail
8. The display indicates “Electrode fault (0x71020000)” Pass Fail
9. The electrotherapy program is resumed on CH2 Pass Fail
10. The display indicates “Electrode fault (0x41020000)” Pass Fail
11. The electrotherapy program is resumed on CH1 Pass Fail
Leakage Current
12. The Voltage DC is between 0mV and 10mV after stabilization. Pass Fail
The Voltage AC is between 1mV and 100 mV after stabilization.
13. The Voltage DC is between 0mV and 10mV after stabilization. Pass Fail
The Voltage AC is between 1mV and 100 mV after stabilization.
14. The Voltage DC is between 0mV and 10mV after stabilization. Pass Fail
The Voltage AC is between 1mV and 100 mV after stabilization.
Battery (optional test)
15. The battery is detected Pass Fail
Visual Check
16. All tests performed are passed and the battery is removed. Pass Fail
Current date (YYYY MM DD): / /
Operator name: _______________________________________
Operator signature: ___________________________________