This device should be used only under the continued supervision of a physician or licensed practitioner.
•
Contaminated sponges, electrodes, leadwires, and gel can lead to infection.
•
Use of electrode on multiple patients can lead to infection.
•
Do not apply electro stimulation treatment during bath, shower, sauna,..
•
DO NOT operate the Intelect® Mobile 2 within the vicinity or environment of an ultrasonic diathermy
system.
•
DO NOT operate the Intelect® Mobile 2 within the vicinity or environment of any
microware and RF shortwave
diathermy system.
•
DO NOT operate this unit in an environment where other devices are being used that
intentionally radiate electromagnetic energy in an unshielded manner.
•
Electronic monitoring equipment (such as ECG monitors and ECG alarms) may not operate
properly when electrical stimulation is in use.
•
Simultaneous connection of a PATIENT to a high frequency surgical ME EQUIPMENT may result in burns
at the site of the STIMULATOR electrodes and possible damage to the STIMULATOR.
•
Portable RF communications equipment should be used no closer than 30 cm (12 inches) to any
part of the Intelect Mobile 2, including cables specified by the manufacturer. Otherwise,
degradation of the performance of this equipment could result.
•
Use of accessories, transducers and cables other than those specified or provided by the manufacturer
of this equipment could result in increased electromagnetic emissions or decreased electromagnetic
immunity of this equipment and result in improper operation.
•
Device is designed to comply with electromagnetic safety standards. This equipment generates,
uses, and can radiate radio frequency energy and, if not installed and used in accordance with
instructions for use, may cause harmful interference to other devices in the vicinity. Harmful
interference to other devices can be determined by turning this equipment on and off. Try to
correct the interference using one or more of the following:
•
Reorient or relocate the receiving device
•
Increase the separation between the equipment
•
Connect the equipment to an outlet on a different circuit from that to which the other device(s) are
connected
•
Consult your authorized DJO dealer for help.
•
Disconnect the system from the power source before attempting any maintenance, installation,
removal, or replacement procedures to prevent electrical shock and possible damage to system.
•
The Intelect® Mobile 2 may be susceptible to Electro-Static Discharge (ESD) at greater than ±4 kV
when first grasping either the Ultrasound or Laser applicator. In the event of such a discharge,
the Intelect® Mobile 2 may experience communication loss. The Intelect® Mobile 2 will
terminate all active outputs (stim, ultrasound,), automatically place the unit in a safe state. TBC,
keep it for the moment. It will depend on formal EMC results
•
To prevent Electro-Static Discharge (ESD) at greater than ±4 kV:.
•
Grasp and hold the Ultrasound prior to starting treatment. If the applicator must be put
down prior to completion of treatment, stop the current treatment first and then place
the applicator in the holder.
•
Maintain humidity in the use environment to at least 50% relative humidity.
•
Floors should be wood, concrete, or ceramic tile. If floors are covered with synthetic
material, DJO recommends implementing additional controls to maintain relative
humidity to at least 50%.
•
Communicate these ESD-precautionary procedures to healthcare staff, contractors, visitors,
and patients. TBC, keep it for the moment. It will depend on formal EMC results