EN
4
| Intelect
®
Mobile 2 - Service Manual 8
Stimulus delivered by the TENS waveforms of this device, in certain configurations, will deliver a charge of 25
microcoulombs (µC) or greater per pulse and may be sufficient to cause electrocution. Electrical current of
this magnitude must not flow through the thorax because it may cause a cardiac arrhythmia. TBC
DO NOT connect the unit to an electrical supply without first verifying that the power supply is the correct
voltage. Incorrect voltage may cause unit damage, malfunction, electrical shock, fire, or personal injury. Your
unit was constructed to operate only on the electrical voltage specified on the Voltage Rating and Serial
Number Plate. Contact your DJO dealer if the unit is not properly rated.
Device is not designed to be used in oxygen rich environment, Explosion hazard if the device is used in the
presence of flammable anesthetic mixture with air, oxygen, or nitrous oxide
DANGER