MANUALPC-2010 CHAPTER IV - OPERATION Page 25
8.1 - INITIAL SETUP SUBMENU
The Initial Setup Submenu is used to specify basic operating
conditions of the CHEMTROL® PC. It is accessed through the
Configuration Menu on the last line of the Display Screen.
8.1.1 - Language
The Language Submenu allows the user to select either one of
three languages: English, French or Spanish for all displays
screens. The standard (default) language is English.
Language changes take effect immediately.
8.1.2 - Units
T
he Units Submenu allows the choice system of U.S. or Metric
units to be used throughout the program. The standard
(default) value is the U.S. system. The change of units takes
place immediately.
Unit equivalencies are as follows:
U.S. METRIC
ORP mV mV
Sanitizer ppm mg/l
pH pH pH
Conductivity µ S µ S
TDS ppm mg/l
Temperature F C
Pressure psi kPa
Flow Rate gpm l/m
Flow Mga m3
8.1.3 - Code Number
T
he Code Number Submenu is used to define different
operator access levels. Code numbers may be required for
access at key points in the program and for remote
communications.
Up to ten Code NUMBERS (of one to five digits each - no
letters) may be entered, along with an associated access level
from one to three. Make sure to select an easy to remember
number, such as a familiar name on a standard telephone
keypad.
The following access levels are available:
Level 1: View only,
Level 2: Calibration,
Level 3: All functions.
To clear an existing Code Number, its access level is set to
zero.
Once a Code Number has been acknowledged, it remains
valid for an hour of continuous operation so that the operator
does not have to re-enter it constantly. If necessary, it can be
changed by returning to the Welcome screen.
Submenu 8.1
Submenu 8.1.1
S
ubmenu 8.1.2
Submenu 8.1.3
English
French
Spanish
U.S.
Metric
Code No. 12345
Access 2
Language
Units
Code No.
Clock
Readings
Data Logging
Model Options