Wound Component EST Scanner 19035/19035-M/19035-S
Quick Start Guide
An option of GP-IB/HANDLER/Temperature Compensation 3 in 1
interface is available for purchase.
Impulse Winding Test (Optional)
This EST Scanner is able to install the optional RS232 interface card
and Impulse Winding Tester for wound component pulse test.
Different STEP can call different main wave data that saved in the
Impulse Winding Tester for multiple continuous tests. It also means
to recall different golden samples.
The EST Scanner has 50 sets of memories and each set contains
up to 20 test steps.
A full function front panel calibration is provided.
DCR Mode Temperature Compensation
Since the wire resistance will change following the temperature, the
resistance measurement needs to add the temperature as one of the
test condition. The function of temperature compensation is to
calculate the resistance at another temperature from the temperature
and resistance already known. It is frequently used to convert the wire
resistance under different temperature.
1.3 Ambient Environment
1. Do not use the meter in a
dusty or vibrating location.
Do not expose it to sunlight
or corrosive gas. Be sure
that the ambient temperature
is 0°C ~ +45°C and that the
relative humidity is between
15% ~ 95%.
2. The scanner has been
carefully designed to reduce
the noise from the AC power
source. However, it should
be used in an environment
with lowest noise as
possible. If noise is
Please keep off the followin
g
s:
Dust
Direct Sunlight
Corrosive
Gas
Vibration
3