CI-154 User’s Manual
Page 7
1.5 Specifications
Four sizes: 0.3, 0.5, 1.0, 5.0 m
Meets requirements of ISO 21501-4 - 10% typical
1.0 x 10
6
particles per cubic foot (10% coincidence)
(35.3 x 10
6
particles per cubic meter)
1.00 cubic feet per minute
(0.028 cubic meter or 28.3 liters per minute)
Electronic, automatic closed loop
3/8” (9.525 mm) I.D., up to 20 ft. (6.1 meters) max. length
7.1 per CM >0.3 µm (95% UCL) per ISO 21501-4
(1 count per five minutes)
Replaceable 9.6 Volts, 12.6 Amp hour Nickel Metal Hydride
Battery operation time: 5.5 to 6.5 hours continuous operation
Time to charge battery:3.5 to 4.0 hours
TFT Color LCD, ¼ VGA touch screen
3,000 samples, date and time stamped
Up to 20 users, besides supervisor, with password
Four security levels with user authentication
Up to 100 per program; 16 characters long.
Up to 30 user-defined and named, containing IDs and settings.
ASCII command set, CSV data output with check code
+5 VDC @ 400 mA provided on pin 9
10/100Base-T, half or full duplex (auto sensing), http web
server, Ethernet Modbus
®
server and TELNET server
Transfer of stored data memory to jump drive
Transfer of user-defined programs to/from jump drive
Transfer of unit settings to/from unit via jump drive
Dry contacts, normally open or closed via alarm connector.
Connector also has +12VDC @ 250mA.
Programmable for all particle channels and flow.
H-W-D: 8.65 x 8.5 x 9.5 inches (21.97 X 21.6 X 24.1 cm)
11.0 lb (4.99 Kg) without battery
14.25 lb (6.46 Kg) with battery
Auto switching: 100–240 VAC, 47/63 Hz, 90VA max
two 5 X 20 mm, 250V/2.0 Amp Slow
0–36º C, 10–90% RH, non-condensing
Factory Prime Calibration complying with ISO 21501-4.
NIST traceable using CalPro 2.00 (or later) software program.
Recommended every 12 months.
Air, nitrogen; consult factory regarding other gases.