Page 54
Sending a request for a number of samples that is greater than the number of
records in memory (e.g., #5,1 with only 5 samples in memory) with result in the
following error message:
EMPTY RECORD
3.3.5 RS-232 and telnet output format
Listed below are the lines that are printed on the fly in response to character
commands. A <CR><LF> is sent after the last character of each line transmitted.
A <BEL> (07H) is sent when the execution of the command is completed. This will
occur for example when the “S” command to start the counter is received and
executed or when the last record of the “A” command is transmitted.
Response to the “A” or “Annnn” command:
Alarm,Date,Time,Location,Unit ID,User ID,Program Name,Run #,Volume,
Count Mode,0.3 Micron,0.5 Micron,1.0 Micron,5.0 Micron, Flow CFM,
Unit S/N,Laser OK,Flow OK,Cal Due
0,04/15/05,15:58:16,LOC ABC,CLIMET 23, ELVIS,PROGRAM1,2,0.2CF,
TOTAL COUNTS,47536,8672,1529,103,1.00,000000,1,1,06/03/09
0,04/15/05,15:58:28,LOC ABC,CLIMET 23, ELVIS,PROGRAM1,2,0.2CF,
TOTAL COUNTS,47525,8669,1534,105, 1.00,000000,1,1,06/03/09
0,04/15/05,15:58:40,LOC ABC,CLIMET 23, ELVIS,PROGRAM1,2,0.2CF,
TOTAL COUNTS,47542,8677,1536,106, 1.00,000000,1,1,06/03/09
Note: Count Mode is the current count mode for RS-232 response. Count Mode is
the mode when data was collected for Telnet response.
Response to the “Y” command:
CI-154 SN000000
Software Revision 1.00
Last Calibration: 06/03/08
Next Calibration: 06/03/09
ID=TEST ID
SAMPLE VOL= 0.2CF
NUMBER OF SAMPLES= 2
INITIAL DELAY=00:00:05
SAMPLE DELAY =00:00:00
4 SAMPLES STORED
UNIT=
USER=
PROGRAM= PROGRAM1
ALTITUDE=0F
PKEY= 285239286233160471096193 (unique number for each computer running the
CalPro software)
CALPRO=02.00 (version 2.00 must be used as a minimum)
USBHOST=Ver 03.68VDAPF (CI-x54 series only.)
CLOCK BATTERY=OK (If main board has coin lithium battery.)