CI-154 User’s Manual
Page 55
Response to the “Y1” command:
CI-154 SN000000
Software Revision 1.00
Last Calibration: 06/03/08
Next Calibration: 06/03/09
ID=TEST ID
SAMPLE VOL= 0.2CF
NUMBER OF SAMPLES= 2
INITIAL DELAY=00:00:05
SAMPLE DELAY =00:00:00
4 SAMPLES STORED
UNIT=
USER=
PROGRAM= PROGRAM1
ALTITUDE=0F
PKEY= 285239286233160471096193 (unique number for each computer running the
CalPro software)
CALPRO=02.00 (version 2.00 must be used as a minimum)
CLOCK BATTERY=OK (If main board has coin lithium battery.)
CH1,DISP=OFF,ALARM=OFF,PRN=OFF
CH2,DISP=ON,ALARM=100,PRN=ON
CH3,DISP=OFF,ALARM=OFF,PRN=OFF
CH4,DISP=ON,ALARM=1,PRN=ON
FLOWHI,ALARM=OFF
FLOWLO,ALARM=OFF
PRN=END OF SAMPLE,PRINT AVG AFTER N=OFF,STOP AFTER N=ON,LARGE=ON
Response to the “y” command:
Current Laser Power = 1177mV
Nominal Laser Power = 1185mV
Current Laser current = 63mA
Nominal Laser current = 65mA
Response to the “F” command:
1,04/15/05,16:00:38, LOC ABC,CLIMET 23, ELVIS, PROGRAM1,2, 0.2CF,TOTAL
COUNTS,47536,8672,1534,103, 1.00, 000000,1,1,06/03/09
End of Sample:
0,04/15/05,16:01:02, LOC ABC,CLIMET 23, ELVIS, PROGRAM1,1, 0.2CF,TOTAL
COUNTS,47352,8911,1567,133, 1.00, 000000,1,1,06/03/09
End of Sample or response to the “F” command with extra information added.
(See command “b!”):
Extra information following cal due date is: time to collect sample volume in 1/100’s
of seconds, 4 channels of count always in total count mode, 32 bit check code.
0,05/20/10,13:48:36, LOC ABC,CLIMET 23, ELVIS, PROGRAM1,1, 0.2CF,TOTAL
COUNTS,47352,8911,1567,133, 1.00, 000000,1,1,06/03/09,375,47352,8911,1567,133,
6C1D71A3

Table of Contents