CI-154 User’s Manual
Page 49
ASCII
Note
FUNCTION
c
Force cumulative count modeas opposed to concentration
mode.
C
Force concentration mode: counts/cubic foot.
H
Force concentration mode: counts/cubic meter.
H
Force concentration mode: counts/liter.
E
Enable touch screen keyboard.
D
Disable touch screen keyboard.
F
Send current date, time, location ID, unit ID, user ID,
program name, run #, volume, count mode, counts, flow,
unit S/N, laser status, flow status and cal due date. (Cal due
date is for CI-x54 series only.)
~
(tilde) Clear command buffer. No <CR> required.
Y
Send model number, serial number, software revision
number, last calibration date, next calibration date, current
location ID, current sample volume, number of samples,
initial delay, and delay between samples, number of stored
samples, Unit ID string, current user name, current program
name, altitude setting, calibration computer product key,
and the version of CalPro that was used during the
calibration of the unit.
Y1
(1)
Same as Y command above with additional: displayed
channels settings, particle alarm settings, printed channels
settings and flow alarm settings.
Y
Send laser power current and nominal values, current and
initial laser current.
B
Send most recent sample completed after power up.
!
Send board # and dash #, mV range of DSP A/D converter,
sea level flow rate, and “calpro_1,idpro_1”.
\
Forces transmitted data to use a “\” delimiter instead of the
default “,” delimiter. This applies to end-of-sample data and
responses to the A, F and ! commands. Cycling power on/off
restores the default “,” delimiter.
?x
(s2)
Transmit the index (location) of the next sample, the
number of samples in memory, the total size of memory and
the date format.
Ex: 0,3,3000,MDY 0 is oldest data index, 3 is next index
that will be written to, 3000 is size of stored data memory
(last index is 2999), date format is month, day, year.
X
(t2)
?X
(s2)
Returns stored data memory data dump header
X
(t2)

Table of Contents