NIKI V4 Operator’s Manual 32
Changing the current default “Pressure” Limit to another
Press SELECT key continuously.
Switch on the pump by using the ON/OFF key.
Press the RUN/HOLD key, PRES shows on display for one
second, thereafter the current pressure will appear.
RATE-
-
VOL.-
By pressing the SELECT key scroll through, and choose the
desired limit.
)
NOTE
The following pressure parameters can be set:
Low = 0.3 bar. (5 psi)
RATE-
-
VOL.-
Normal = 0.5 bar. (7 psi)
RATE-
-
VOL.-
High = 0.7 bar. (10 psi)
RATE-
-
VOL.-
)
NOTE
The selected pressure setting will remain in the pump’s memory until
changed again.
)
Press the RUN/HOLD key, in order to store the parameter in the pump’s memory.
NOTE The pump has an antibolus system, which prevent a bolus delivery
to the patient, once a downstream occlusion is released.