CME BodyGuard 323 Service Manual

CME BodyGuard 323
50 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Page #1 background imageLoading...
Page #1 background image
Service Manual
for
BodyGuard 323, 545, & 575
Infusion pumps
0344
Manufacturer:
Caesarea Medical Electronics Ltd.
International Toll Free: +800-323-575-00
European address:
Staufenburgstr. 23 Lichtenstein
P.o.b. 1248 Lichtenstein 72805 Germany
PN 100-091X ver. 191009
E-mail: sales@cme-infusion.com
www.cme-infusion.com
Question and Answer IconNeed help?

Do you have a question about the CME BodyGuard 323 and is the answer not in the manual?

CME BodyGuard 323 Specifications

General IconGeneral
ManufacturerCME
ModelBodyGuard 323
Operating Temperature5°C to 40°C
Power SupplyAC/DC, Battery
Noise Level< 45 dB

Summary

Introduction

Warnings and Cautions

Service Process Diagram

Customer and Device Information & Pre-Service Checklist

Tools and Test Equipment

Adjustable Operating Parameters and Options

BodyGuard 323 Technician Menu

Lists parameters adjustable via the Technician Menu for BodyGuard 323.

Change Set Up Menu

Program Lock

Keypad Lock

BodyGuard 545 and 575 Operation Modes

Technician Menu for 545/575

Details parameters adjustable via the Technician Menu for 545/575 models.

Level 1 Menu

Operational Checklist and Performance Acceptance Test

Main Self Test

Comprehensive self-test for keypad, display, alarms, door, air, motor, pressure.

Open Door Alert Function

Procedure to test the pump's door open alert functionality.

Air Detect Function

Test for the pump's air bubble detection system.

Battery Operation Test

Procedure to test the pump's battery performance and life.

Restore Software Default Settings

Guide for restoring the pump's software to factory default settings.

Pole Mount Charger Voltage Test

Procedure for testing the voltage output of the pole mount charger.

Verify DC Extension Cable Connector Operation

Tests the functionality of the DC extension cable connector.

Volume Delivery

Procedure to assess and calibrate the pump's volume delivery accuracy.

Occlusion Pressure

Procedure to assess and calibrate the pump's occlusion pressure detection.

Ground Strap Test

Checks continuity for ground strap connection.

QC and Calibration Labels

Process for applying QC and calibration labels after service.

Calibration

Volume Calibration

Detailed procedure for calibrating the pump's volume delivery accuracy.

Occlusion Pressure Calibration

Troubleshooting

Service actions

Disassembly / Assembly of Pole Mount Charger

Procedures for disassembling and assembling the pole mount charger.

Disassembly / Assembly of Pump

Steps for taking apart and reassembling the BodyGuard infusion pump.

Replacement of Pressure Sensor

Guide for replacing the pump's pressure sensor component.

Replacement of Air Sensors

Instructions for replacing the air sensor(s) in the pump.

Replacement of Pump Motor

Steps for replacing the motor within the infusion pump.

Replacement of LCD

Guide for replacing the pump's Liquid Crystal Display.

Software Update and Replacement

Process for updating or replacing the pump's software.

Restore Software Default Settings

Service Center Recommended Parts

Cleaning and Storage

APPENDIX A

Related product manuals