Occlusion Pressure Calibration
Background:
Occlusion pressure calibration will allow the pressure alerts to activate
correctly. The calibration is accomplished by adjusting the pump to display
the correct pressure at two different pressure levels. This will enable the
pump to display the pressure accurately anywhere within the operational
range to the specification tolerance. Perform the Occlusion Pressure
Calibration using the test set-up diagramed below.
Acceptance Criteria:
Alarm should occur at 10+4/-2psi for BodyGuard 575 and 14+4/- 2 for the
BodyGuard 323, & 545.
Required Equipment:
Pressure Gauge (.1 psi resolution)
BodyGuard tube set
Collection vessel
IV bag
BodyGuard pole mount charger and power cord
IV pole or similar mounting surface
BodyGuard Pump