1.4 Pulse Oximeter Charger
1. Plug in Pole Mount Charger with Pulse Oximeter connector.
2. Attach working pump to charger.
3. Enable Pulse Oximeter feature through Change Set Up Menu
4. Note on service record, if Pulse Oximeter feature was not enabled.
5. If it was enabled, attach working Pulse Oximeter to connector.
6. Activate Pulse Oximeter by starting program.
7. If it does not work replace Charger.
8. Perform Operational Checklist and Performance Acceptance Test