of 5047
8. Restore Software Default Settings
Note: Software default settings should only be restored before beginning the
last Operational Checklist and Performance Acceptance Test after service has
been completed.
Note: Complaint pumps will have software default settings restored after
notification of complaint closure.
1. Power the pump on while pressing the INFO key to enter the Technician Menu.
2. Enter the appropriate Technician Code for the device.
3. Use the arrow keys to scroll to the option “Software Factory Setting”.
4. Press the START/OK key.
5. Press the START/OK key to reset the software factory settings to default.
6. Select the “Restart Pump” option.
7. Press the START/OK key.
8. Verify factory default settings have been reset to default values in chart below.
To assure default settings are restored for BodyGuard 545, and 575, check that
protocol parameters in the “Select Protocol” option have been erased.
9. Add “Software Default Settings Restored” insert to pump packaging. This
notifies customer that any desired customer programmed setting must be
reprogrammed prior to use.
10. Perform Operational Checklist and Performance Acceptance Test
BodyGuard Pump Software Default Settings
BG 323 BG 545 BG 575
Settings Default
Value
Setting Default
Value
Setting Default
Value
Pressure default 14 psi Pulse Ox Enabled Pulse Ox Enabled
Program Continuous Key stroke Volume Lowest setting Key stroke Volume Lowest setting
Priming rate Not adjustable Occlusion Pressure High 21 PSI Occlusion Pressure High 10 PSI
Max bolus volume 100 ml Priming Volume 5 ml Priming Volume 5 ml
Titration bolus On Priming Rate 400 ml/hr Priming Rate 400 ml/hr
Bolus rate 600 ml/hr Maximum CA Bolus 5 ml Maximum CA Bolus 5 ml
Maximal rate 1200 ml/hr Default Bolus Rate 0 ml/hr Default Bolus Rate 0 ml/hr
KVO rate 0.0 ml/hr Maximal Basal Rate 30 ml/hr Maximal Basal Rate 5 ml/hr
Flow control 0 drops/ml KVO Rate 0 ml/hr KVO Rate 0 ml/hr
Set loading test Off Program Limits OFF, volume Program Limits OFF, volume
Air bubble 0.3 ml Air In Line Limit 0.5 ml Air In Line Limit 0.5 ml
Store last program Off LED Indicator ON LED Indicator ON
Operation led On Bolus Key OFF Bolus Key OFF
Time option Off Titration Option OFF Titration Option OFF
Delay before start Off