Warnings and Cautions
Warning: Warnings advise you of circumstances that could result in
injury or death to the patient or operator. Read and understand this
manual and the Operation’s Manual taking note of all warnings before
operating or performing service on the BodyGuard Infusion System.
Caution: Cautions advise you of circumstances that could result in
damage to the device. Read and understand this manual and the
Operation’s Manual taking note of all cautions before operating or
performing service on the BodyGuard Infusion System.
Note: Indicates that the information that follows is additional important
information or a tip that will help you when operating or performing
service on the BodyGuard Infusion System.
Warnings
To avoid possible personal injury or loss of life, observe the following:
Read the entire Operation’s Manual before using the pump, since the text
includes important precautions.
The maximum volume that may be infused under SINGLE FAULT
CONDITION is 0.1 ml.
Voltage present on internal components may cause severe shock or death
upon contact. Disconnect the charger from AC power prior to opening the
casing. Only authorized service personnel should open the pump cover.
Blown fuses could cause a fire hazard. Replace blown fuses on the
charger only with fuses of the same type and rating (see fuse values on
the charger PCB).
The equipment is not suitable for use in the presence of flammable
anesthetic-air/oxygen/nitrous oxide mixture. Do not use the system in the
presence of these gases.
Make sure the pump is attached securely to the charger, which is
connected snugly to an IV pole.
A kinked or occluded IV line may impair the operation of the pump and
the accuracy of the infusion. Before operation, verify that the IV line is
not kinked or occluded.
The BodyGuard should be operated only with BodyGuard tubing sets.
Use of administration sets other than manufacturer-produced BodyGuard
tubing set may impair the operation of the pump and the accuracy of an
infusion.