Change Set Up Menu
To access Change Set Up Menu:
1. Turn the pump on in the normal operation mode.
2. To access the Change Set Up Menu, press the INFO key multiple times until
the “Change Set Up” option is displayed,(it is always displayed as the 4th
option).
3. Select “Change Set Up”, press OK.
4. Scroll down to the option “More”, press OK.
5. Enter access “Change Set Up” Code, press OK.
The Change Set Up Menu is intended for use by clinical personnel and authorized
service technicians.
Item
Parameter Description
1 restart pump Allows clinical personnel to return pump to normal operation mode.
2 pressure default Allows clinical personnel to set the default occlusion alarm pressure.
The pump will default to this pressure setting each time the pump is
started.
3 select program Allows clinical personnel to select program from list (Continuous,
Intermitent, PCA, TPN, or 25 steps).
4 priming rate Allows clinical personnel to set priming rate.
5 max bolus volume Allows clinical personnel to set maximum bolus volume (0-100 ml).
6 titration bolus Allows clinical personnel to set titration bolus (On/ Off).
7 bolus rate Allows clinical personnel to set bolus rate (0-1200 ml/h).
8 maximal rate Allows clinical personnel to set maximal rate (0.1-1200 ml/h).
9 KVO rate Allows clinical personnel to set KVO rate (0.0-5.0 ml/h).
10 flow control Allows clinical personnel to set flow control (0 - 20 drops/ml).
11 set loading test Allows clinical personnel to activate set loading test (On/Off).
12 air bubble size Allows clinical personnel to set air bubble size (0.1 - 2.0 ml).
13 store last program Allows clinical personnel to store last program (On/Off).
14 operation LED Allows clinical personnel to set operational indicator LED (On/Off)
15 time option Allows clinical personnel to set time option (On/Off).
16 delay before start Allows clinical personnel to set delay before start (On/Off).
17 language Allows clinical personnel to choose language (English, Spanish).
18 set time and date Allows clinical personnel to date and time (dd, mm, yy, hh, mm).
19 history Allows clinical personnel to review event history settings and alarms
(for last 500 events)