Program Lock
To access Program Lock:
1. Turn the pump on in the normal operation mode.
2. To access the Change Set Up Menu, press the INFO key multiple times until
the “Change Set Up” option is displayed,( it is always displayed as the 4
th
option).
3. Select “Change Set Up”, press OK.
4. Scroll down to the option “More”, press OK.
5. Enter Program Lock Code, press OK.
6. Select ON to activate lock (Off/On).
The Program Lock is intended for use by clinical personnel and authorized service
techncians.
Item
Parameter Description
1 Locks ability to
change programs
Enables clinical personnel to lock out the pump so that programs can
only be started or stopped.
Keypad Lock
To access Keypad Lock:
1. With the pump operating in normal operation mode, to activate the Keypad
Lock: Press and Hold the INFO Key until the bar graph is filled.
2. Repeat step 1 to deactivate.
The Keypad Lock is intended for use by clinical personnel and authorized service
technicians.
Item
Parameter Description
1 Locks all keypad
buttons except the
INFO START/OK
and STOP keys
Enables clinical personnel to lock all buttons but the INFO,
START/OK and STOP keys to prevent tampering with the pump
during normal operation.