Calibration
Volume Calibration
Background:
Volume calibration is performed to enable the pump to deliver
accurately.
Perform the volume calibration using the test set-up diagramed below.
Acceptance Criteria:
The pump must be within +/- 5% to pass this test.
Required Equipment:
Scale with 0.01 gm degree accuracy
BodyGuard tube set
Collection vessel
IV bag
BodyGuard pole mount charger and power cord
IV pole or similar mounting surface
BodyGuard Pump