2.3 Replacement of Pump Door Magnet
1. Replace missing door magnet by securing new magnet into recess in top of
door with press fit or glue.
2. Assure that magnet triggers door open alert and does not interfere with door
operation.
3. If service has been completed, perform Operational Checklist and
Performance Acceptance Test.
2.4 Replacement of Pump Door or Replacement of Pressing Plate
1. Use a 2.5 mm hex key to remove the security
grub screws (M2 x 5) at both ends of the hinge
(as shown on the diagram).
2. Using a Phillips head screw driver, remove
the door-mounted air sensor, fixed with one
M2 screw. Pull the sensor out from its
recess.
Door Air Sensor
2.5 mm security grub screws
3. Lift the air sensor cable and carefully using a 2 mm shaft, push the hinge
pin from the air sensor side (left side) and remove it from the pressure
sensor side. THE HINGE PIN WILL ONLY GO ONE WAY. The hinge
pin must be removed from the right side of the pump and reinserted into
the right side of the pump. Pushing the hinge pin the wrong way will
cause damage to the pump.
4. Removing the hinge pin allows for separation and removal of the door, pressing
plate, and springs.