BodyGuard 545 and BodyGuard 575
The BodyGuard 545 and 575 pumps have adjustable operating parameters. These
operating parameters and options may be viewed and adjusted to modify the
operation of the pump. The operating parameters and options available for the
BodyGuard 545, and 575 are described in the Technician Menu, Level 1 Menu and
Keypad Lock.
Technician Menu
To access the Technician Menu, press and hold the INFO key during the power on
operation until ‘Technician Code’ prompt appears. The display will show the software
version for 2 seconds before the access code prompt. Enter Technician Code to access
the menu.
The pump will display all the parameters that can be set, calibrated or tested. The
technician can scroll through all parameters using the ARROW keys (2 & 0 keys on
the pump).
Item
Parameter Description
1
Restart Pump Exit Technician Mode.
2
Serial Number Displays serial number & production date.
3
Manual Calibration Summary of volume & pressure calibrations and ability to check/re-
enter software default calibrations after service.
4
Main Self-Test Performs keypad, display, alarm sound, door sensor, air sensor,
temperature, motor & encoder, and pressure sensor tests.
5
Volume Test
Program rate & volume to be delivered during flow rate test.
6
Manual Test Same as Main Self-Test but with a menu so technician can focus on
individual tests.
7
Pressure Calibration Allows manual calibration of pressure settings.
8 Volume Calibration Allows manual calibration of pumping mechanism.
9
Software default
Setting
Caution: pressing START/OK will restore software default defaults
and delete all pre-set protocols and set-up changes.
10
Reset Calibrations Caution: pressing START/OK restores default settings. Pump must
be re-calibrated after reseting calibrations.
11
Operating hours Hours from last service – reset to zero by pressing STOP/NO key.
12
Access Codes Displays current access codes and allows change.