Drugs must not be administered to the epidural space unless the drugs are
indicated for this purpose and are administered in accordance with the
indications included in the manufacturer’s package insert accompanying
the drugs.
Epidural administration of drugs other than those indicated for epidural
use could result in serious injury to the patient.
Any adjustments, maintenance, or repair of the uncovered pump may
impair the operation of the BodyGuard Infusion System and/or the
accuracy of the infusion. Only CME authorized technicians should
perform any adjustments, maintenance, or repair of the uncovered pump.
Any adjustments, maintenance, or repair of the uncovered pump while
connected to the power should be avoided.
The BodyGuard Infusion System should be operated within a temperature
range of 15°C (50°F) to 45°C (115°F) and at up to 85% humidity.
Operating the pump at temperatures and/or humidity other than within
this range may affect accuracy.
Unsafe operation may result from using improper accessories. Use only
accessories and options designed for this system and supplied by
manufacturer.
The BodyGuard tubing sets should not be use for blood, blood products or
nitroglycerin administration.
Battery charging is enabled as long as the charger cord is connected to
AC power and the pump is in the charger. Switching the pump off does
not disconnect it from AC power. To disconnect from AC power, remove
the charger cord from AC power. To disconnect pump from AC power,
remove it from the charger.
Dropping the BodyGuard Infusion System could cause damage to
components. If the pump is dropped, return the pump for inspection by
qualified service personnel.
Use aseptic technique. Patient infection may result from the use of non-
sterile components. Maintain sterility of all disposable components and
do not re-use single use IV sets.
When operating the pump in PCA mode with a rate of 0.0 ml/hr there is a
hazard of blood clot formation. Connect saline infusion in parallel to
avoid this problem.
Do not operate the pump near high-energy radio-frequency emitting
equipment, such as electro-surgical cauterizing equipment. False alarm
signals may occur.
Watch your fingers / nails when opening and closing the pump door.