1. Place the BodyGuard unit face down on the electro static discharge mat.
Ensure that the ESD mat is free of debris and other objects that could cause
scratches to the pump housing.
2. Using the #0 x 2-½ Phillips screwdriver remove the four screws from the back
of the unit. Two of the screws are located at the top of the unit underneath the
battery pack. The remaining screws are located at the bottom of the unit
underneath the quality check label.
3. After removing the four screws from the back of the unit gently separate the
back portion of the case from the front portion of the case. Set the back portion
of the case aside and remove the screws from the case. Set the screws aside for
later use.
4. Insert the battery into back portion of the case.
5. Using the five pin male to male wire, connect the PCB to the BodyGuard
programming board by connecting the five pin connector wire to pin JP1.
6. Connect the RS-232 cable to the BodyGuard programming board and the com
port on the back of the computer.
7. Carefully place the rear portion of the case onto the front portion of the case
without disturbing the five pin wire connection.
8. Open the Prog08sz software. If the setup is connected properly the pump will
begin to beep shortly after the programming software is opened.
9. Select the button labeled “Close COM Port”.
10. Set the correct port for the RS-232 cable by using to drop down menu labeled
“Port”.
11. Change the baud rate to 8861 by using the drop down menu labeled “Baud”.
12. Check the box next to “IGNORE security failure and enter monitor mode”.
13. Select “Contact target with these settings”.
14. Select the program algorithm “908_az60a_highspeed.08p”. Make sure to select
the program algorithm that was saved as a file not the algorithm that is included
with the programming software.
15. Specify the S record by selecting it from the pick window. Use the proper file
(i.e. USA35E13.s19).
16. Erase the module by selecting EM from the pick window. After the module has
been erased reconnect with security.
17. Program the module by selecting PM from the pick window.
18. Select Quit from the pick window.
19. Disconnect the pump from the programming board.
20. Reassemble the pump and recalibrate.
21. Perform Operational Checklist and Performance Acceptance Test