EasyManuals Logo

CME NIKI V4 User Manual

CME NIKI V4
54 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #35 background imageLoading...
Page #35 background image
NIKI V4 Operator’s Manual 35
Failure Identification
Specific errors which may occur in the operation of the NIKI V4 Volumetric Infusion
Pump are represented on Section 3 pg.25-27. On the biomedical mode. The user can
display the last 10 failures which are stored in the device's memory.
The failure information, displayed, can be used to determine the nature of a device
failure and to troubleshoot its cause.
Go to biomedical code. The display will show the last 10 errors, each press on Run/Hold
mode key will display another error, scrolling from the last to the 10
th
.
Refer to Alarm condition pg.28 for descriptions of all failures possible and instructions
on how to correct each failure.
Disassemble / Reassemble and calibration procedures are located in section 4.5.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the CME NIKI V4 and is the answer not in the manual?

CME NIKI V4 Specifications

General IconGeneral
BrandCME
ModelNIKI V4
CategoryMedical Equipment
LanguageEnglish

Related product manuals