NIKI V4 Operator’s Manual 35
Failure Identification
Specific errors which may occur in the operation of the NIKI V4 Volumetric Infusion
Pump are represented on Section 3 pg.25-27. On the biomedical mode. The user can
display the last 10 failures which are stored in the device's memory.
The failure information, displayed, can be used to determine the nature of a device
failure and to troubleshoot its cause.
Go to biomedical code. The display will show the last 10 errors, each press on Run/Hold
mode key will display another error, scrolling from the last to the 10
th
.
Refer to Alarm condition pg.28 for descriptions of all failures possible and instructions
on how to correct each failure.
Disassemble / Reassemble and calibration procedures are located in section 4.5.