NIKI V4 Operator’s Manual 39
Air in Line Alarm Test
1. Press on the door’s latch to open pump’s door. Insert an
administration set (which the pump has been calibrated to)
in the pump’s tubing guide.
press here
2. Close the door. Press on the door recessions until a click is
heard.
3. Set the infusion rate to 500ml/hr. Use the arrows key.
RATE-
-
V
OL.-
.
4. Press the RUN/HOLD key to start operation.
Turn the drip chamber, up side down, to allow 5 mm of
air to enter into the administration set.
5. AIR alarm will activate as soon as the air bubble enter the
set segment located behind the pumping door.
RATE-
-
V
OL.-
7. Press the RUN/HOLD key to turn the alarm off.
8. HOLD appears on the display.
RATE-
-
V
OL.-